Properties

Protocol PDF Click for Protocol
Catalog Number TK03691
Species Human
Design Sandwich
Standards 7 standards, serially diluted from 1 prepared lyophilized standard.
Controls
Sample Types Serum, plasma and other biological fluids
Sample Volume 100 μL
Assay Desc.
Standard Range 62.50—4000 pg/mL
Sensitivity 37.50 pg/mL
Radioactivity
Other Names SPPase1
Storage 2 - 8°C
Postscript For research use only, not for use in diagnostic procedures.

Intended use

This ELISA kit applies to the in vitro quantitative determination of Human SGPP1(Sphingosine-1-Phosphate Phosphatase 1) concentrations in serum, plasma and other biological fluids.

Test principle

This ELISA kit uses the Sandwich-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with an antigen specific to Human SGPP1(Sphingosine-1-Phosphate Phosphatase 1). Standards or samples are added to the micro ELISA plate wells and combined with the specific antigen. Then a biotinylated detection antigen specific for Human SGPP1(Sphingosine-1-Phosphate Phosphatase 1) and Avidin-Horseradish Peroxidase (HRP) conjugate are added successively to each micro plate well and incubated. Free components are washed away. The substrate solution is added to each well. Only those wells that contain Human SGPP1(Sphingosine-1-Phosphate Phosphatase 1), biotinylated detection antigen and Avidin-HRP conjugate will appear blue in color. The enzyme-substrate reaction is terminated by the addition of stop solution and the color turns yellow. The optical density (OD) is measured spectrophotometrically at a wavelength of 450 nm ± 2 nm. The OD value is proportional to the concentration of Human SGPP1(Sphingosine-1-Phosphate Phosphatase 1). You can calculate the concentration of Human SGPP1(Sphingosine-1-Phosphate Phosphatase 1) in the samples by comparing the OD of the samples to the standard curve.

Typical data

As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, washing technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data is provided below for reference only.

Concentration(pg/ml) O.D Average
4000 2.31
2.342
2.326
2000 1.545
1.565
1.555
1000 0.88
0.843
0.862
500 0.443
0.475
0.459
250 0.239
0.231
0.235
125 0.184
0.16
0.172
62.5 0.121
0.13
0.126
0 0.074
0.076
0.075

Precision

Intra-assay Precision (Precision within an assay):
3 samples with low, middle,high concentration of Human SGPP1(Sphingosine-1-Phosphate Phosphatase 1). And the sample of Human SGPP1(Sphingosine-1-Phosphate Phosphatase 1) were tested 30 times on one plate, respectively.
Inter-assay Precision (Precision between assays):
3 samples with low, middle,high concentration of Human SGPP1(Sphingosine-1-Phosphate Phosphatase 1). And the sample of Human SGPP1(Sphingosine-1-Phosphate Phosphatase 1) were tested on 3 different plates, 30 replicates in each plate.

  Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
n 30 30 30 30 30 30
Mean (pg/ml) 195 618.75 1507.5 187.5 558.75 1436.25
Standard deviation 10.11 27.94 61.6 9.98 25.19 57.83
C V (%) 5.18 4.52 4.09 5.32 4.51 4.03

Recovery

Matrices listed below were spiked with certain level of Human SGPP1(Sphingosine-1-Phosphate Phosphatase 1) and the recovery rates were calculated by comparing the measured value to the expected amount of Human SGPP1(Sphingosine-1-Phosphate Phosphatase 1) in samples.

Matrix Recovery range (%) Average Recovery (%)
Serum (n=5) 92-104 96
EDTA plasma (n=5) 88-101 92
heparin plasma (n=5) 81-102 94

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Human SGPP1(Sphingosine-1-Phosphate Phosphatase 1) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

    Serum (n=5) EDTA plasma (n=5) heparin plasma (n=5)
1:2 Range (%) 87-108 89-106 91-101
Average (%) 95 94 94
1:4 Range (%) 90-104 88-102 87-108
Average (%) 93 93 94
1:8 Range (%) 92-100 94-106 93-104
Average (%) 93 93 93
1:16 Range (%) 88-101 87-105 89-99
Average (%) 95 93 94

Stability

The stability of Human SGPP1(Sphingosine-1-Phosphate Phosphatase 1) ELISA Kit is determined by the loss rate of activity. The loss rate of Human SGPP1(Sphingosine-1-Phosphate Phosphatase 1) ELISA Kit is less than 5% within the expiration date under appropriate storage condition. To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Item Specifications Storage
Micro ELISA Plate(Dismountable) 8 wells ×12 strips -20℃, 6 months
Reference Standard 2 vials
Concentrated Biotinylated Detection Ab (100×) 1 vial, 120 μL
Concentrated HRP Conjugate (100×) 1 vial, 120 μL -20℃(shading light), 6 months
Reference Standard & Sample Diluent 1 vial, 20 mL 4℃, 6 months
Biotinylated Detection Ab Diluent 1 vial, 14 mL
HRP Conjugate Diluent 1 vial, 14 mL
Concentrated Wash Buffer (25×) 1 vial, 30 mL
Substrate Reagent 1 vial, 10 mL 4℃(shading light)
Stop Solution 1 vial, 10 mL 4℃
Plate Sealer 5 pieces
Product Description 1 copy
Certificate of Analysis 1 copy

Other supplies required

Microplate reader with 450 nm wavelength filter
High-precision transfer pipette, EP tubes and disposable pipette tips
Incubator capable of maintaining 37℃
Deionized or distilled water
Absorbent paper
Loading slot for Wash Buffer

Assay procedure summary

1. Prepare all reagents, samples and standards;
2. Add 100µL standard or sample to each well. Incubate 1 hours at 37°C;
3. Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C;
4. Aspirate and wash 3 times;
5. Add 100µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
6. Aspirate and wash 5 times;
7. Add 90µL Substrate Solution. Incubate 10-20 minutes at 37°C;
8. Add 50µL Stop Solution. Read at 450nm immediately.